[Matrix-F] channel is disabled

Submitted by dqhung on 28 June 2017 - 8:00am
Forums:
Hi,
I am using Matrix-F (NIR) of Brucker with 5 channel : A, B, C ,D and E.
But, i just receive result from channel A & B. Channel C, D,E gives me wrong data. So, i checked the signal and spectrum of each channel :
+ A & B : is normal
+ C & D : the signal is weak, i have to zoom if i want to see the spectrum.
+ E : signal line isn't smooth.
After that, I ran each channel individually -> i have the result from five channel, rught result.
I tried to clean probe but notthing change
I hope somebody can help me !!!
Thanks you very much
Hung
muisaranwong
28 Jun 2017
Hello Mr. Hung,
Greetings from Bruker Optics South East Asia. I am taking care of all technical prospect of our customers in SEA and Japan. As your system has been taking care by our Korea team, I already forward your info to them. Would you please send me your contact info by email, so we can get in touch with you either by phone or by email.
Best,
Mui
[email protected]
Permalink
dqhung
28 Jun 2017
Hi Ms. Mui,
Thanks for your help.
I just want try to fix it before contact to Brucker. We already have contact of Mr Jeapil Lee - Brucker in Korean :)
Best Regrad
Hung
Permalink
muisaranwong
28 Jun 2017
Hi,
In process environment, there could be several reasons that cause the weaken of signal such as probe contamination, loosen sma connection, etc.
I think the best way is that our Korean colleague contacted you and perform remote diagnostic.
If you dont here from Mr. JP Lee within 24 hours, please drop me an email. I can help pushing (as I am now in Korea as well).
Best,
Mui
Permalink
dqhung
29 Jun 2017
Hi
Thanks Ms. Mui,
I discussed with him several time before, he is very nice . I just want study a little before ask something.
I will contact to him
Best,
Hung
Permalink